Convenient methods for ingestion of molecular hydrogen: drinking, injection, and inhalation
Kurokawa R, Seo T, Sato B, Hirano S, Sato F. · Medical Gas Research. 2015;5:13.
Study at a glance
Researchers measured how experimental devices produced and retained hydrogen in water, sealed fluids and a gas stream. Two pressurized water systems reached roughly 7–10 mg/L, with losses after opening. A gas device maintained about 3% hydrogen output for three hours. These were laboratory concentration measurements: no people or animals received treatment in this report. The study does not establish health benefits, medical sterility, safe home preparation or the performance of other devices. All authors worked for the company developing the tested apparatuses.
Preclinical
Bench concentration and storage measurements in water, saline and gas systems; no treated people or animals.
Two pressurized hydrogen-water preparation systems; permeation of hydrogen into sealed fluid containers; separate sealed storage approach; electrolysis-derived hydrogen diluted into air. These are experimental preparation methods, not approved clinical directions or consumer-device recommendations. · Two pressurized hydrogen-water preparation systems; permeation of hydrogen into sealed fluid containers; separate sealed storage approach; electrolysis-derived hydrogen diluted into air. These are experimental preparation methods, not approved clinical directions or consumer-device recommendations.
Water concentration measured over preparation and up to seven days sealed / hours after opening; saline storage up to 12 months; gas concentration stability up to three hours.
Two sealed-water systems reached 7.13 ± 0.22 and 10.08 ± 0.34 mg/L after 24 hours; after seven days unopened, reported concentrations were 8.30 ± 0.98 and 10.10 ± 0.21 mg/L. Concentrations declined after opening. These values describe these pressurized experimental systems, not all marketed hydrogen-water bottles. Hydrogen permeation into infusion containers depended on material, thickness and fluid. Approximately 1.0 mg/L was obtained after ten hours immersion; after removal, concentration fell to approximately 0.6 mg/L within one hour in the described condition. Some fluids/containers admitted hydrogen poorly. Concentrations in separately packaged sealed containers were retained over 12 months in individual measurements; this did not establish maintained sterility, chemical purity, clinical shelf life or regulatory approval. Gas output was 2.91 ± 0.08% hydrogen after 30 seconds and remained approximately 3% for three hours. No significant time-point difference was reported. This is device-output stability, not measured inspired concentration or human safety evidence.
Small laboratory measurement series; several fluid measurements are single observations with limited reproducibility assessment.
What kind of evidence is this?
Preclinical
Laboratory comparison of hydrogen preparation and storage methods. Publication carries a Study Protocol heading but reports original experimental concentration measurements; not a randomized efficacy trial or pharmacokinetic animal experiment.
Other molecular hydrogen research
Other forms
Bench concentration/stability measurements; no clinical outcomes
Hydrogen concentration in water/fluids and gas under the specific experimental conditions; not symptom improvement, human tissue exposure or comprehensive device safety.
Methods
Bench concentration and storage measurements in water, saline and gas systems; no treated people or animals.
No treated human participants or animals. Water experiments generally 3–5 independent measurements; gas stability n=3; saline/container results are individual measurements. Do not count cited earlier clinical studies as participants in this report.
Water concentration measured over preparation and up to seven days sealed / hours after opening; saline storage up to 12 months; gas concentration stability up to three hours.
Two pressurized hydrogen-water preparation systems; permeation of hydrogen into sealed fluid containers; separate sealed storage approach; electrolysis-derived hydrogen diluted into air. These are experimental preparation methods, not approved clinical directions or consumer-device recommendations.
Experimental preparation and concentration/stability measurements across water, fluids and gas; no treatment administered.
Two pressurized hydrogen-water preparation systems; permeation of hydrogen into sealed fluid containers; separate sealed storage approach; electrolysis-derived hydrogen diluted into air. These are experimental preparation methods, not approved clinical directions or consumer-device recommendations.
Temperature, preparation time, opening/storage duration and container/fluid type comparisons; no patient placebo arm or disease model.
Outcomes and reported result
Hydrogen concentration in water/fluids and gas under the specific experimental conditions; not symptom improvement, human tissue exposure or comprehensive device safety.
Two sealed-water systems reached 7.13 ± 0.22 and 10.08 ± 0.34 mg/L after 24 hours; after seven days unopened, reported concentrations were 8.30 ± 0.98 and 10.10 ± 0.21 mg/L. Concentrations declined after opening. These values describe these pressurized experimental systems, not all marketed hydrogen-water bottles. Hydrogen permeation into infusion containers depended on material, thickness and fluid. Approximately 1.0 mg/L was obtained after ten hours immersion; after removal, concentration fell to approximately 0.6 mg/L within one hour in the described condition. Some fluids/containers admitted hydrogen poorly. Concentrations in separately packaged sealed containers were retained over 12 months in individual measurements; this did not establish maintained sterility, chemical purity, clinical shelf life or regulatory approval. Gas output was 2.91 ± 0.08% hydrogen after 30 seconds and remained approximately 3% for three hours. No significant time-point difference was reported. This is device-output stability, not measured inspired concentration or human safety evidence.
Complete main article, seven figure captions, container table and disclosures reviewed. No separate supplement identified. Figures were read through captions/text, not independently digitized or visually audited. No separate supplement identified in the reviewed main article. Null findings, discrepancies and unreported values are retained; no independent reanalysis.
Not applicable as a clinical adverse-event assessment; no clinical safety conclusion supported.
A reported association, difference or mechanism is not automatically a clinical benefit.
Limitations and applicability
Small laboratory measurement series; several fluid measurements are single observations with limited reproducibility assessment. No treated participants or animals, clinical efficacy outcomes, actual inspired dose, tissue pharmacokinetics, broad contaminant testing or comprehensive safety assessment. All authors were affiliated with MiZ Co., Ltd. and developed the devices; authors nevertheless declared no competing interests. Preserve both facts. Hydrogen concentration retention does not establish validated medical sterility, approved storage period or suitability for self-prepared infusions. The article calls below-4% gas safe relative to a stated detonation limit; this is not a universal clinical safety threshold. Do not present the preparation recipes as home-use instructions. Figure findings checked against textual captions/results; plots were not independently digitized or visually audited in this review.
The reported population or experimental model and protocol only; clinical benefit and general safety are not established.
No explicit funding statement identified.
Authors declared no competing interests; all author affiliations were MiZ Co., Ltd., and authors developed/prepared the tested apparatuses.
No single-study GRADE certainty rating is assigned. Read how records and evidence assessments are prepared.
Sources and record status
PMID: 26504515 · DOI: 10.1186/s13618-015-0034-2
Published main full text available through the linked original source.
Editorial integration of the supplied source review; not independent medical or statistical validation.
Complete cached official full main XML including introduction, methods, all results and seven figure captions/container table, conclusion, disclosures and 18 references read. No supplementary material identified; no independent plot/image or raw-data audit. Supplementary scope: No separate supplement identified in the reviewed main article.
Complete published main full-text article reviewed using the supplied study-specific packet. Supplementary scope is stated separately; no independent medical, statistical or image-forensic validation.
Ethics and consent: No human/animal administration reported in the article; no clinical ethics approval identified or inferred. Related-data limitations: No participant/animal cohort in this technical report. Cited human disease studies and the separate rat-tissue study PMID 24975958 are background, not new independent data here. Do not attribute those efficacy/PK outcomes to this article.