Respiratory-physiology modeling of therapeutic hydrogen inhalation: defining the fraction of inspired hydrogen (FiH2) and flow-rate requirements.
LeBaron TW, Ohno K, Salomez-Ihl C, Cinquin P, Boucher F, Sano M, Kheir JN. · Respiratory Research.2026;27:259.
Study at a glance
This modelling study examined why the hydrogen flow shown on a device does not tell us exactly how much hydrogen a person breathes in through a nasal cannula. The calculations varied breathing speed, time spent inhaling, nasal capture and possible gas storage in the upper airway. The same device flow could produce very different inspired hydrogen fractions under different assumptions. The paper recommends reporting the hydrogen fraction in inspired air rather than device flow alone. Its numerical estimates and suggested target concentrations are model outputs and the authors’ interpretation of earlier research. They were not prospectively validated in a new group of patients and do not establish a universal effective or safe dose. An average calculated concentration below the hydrogen flammability limit also does not rule out locally flammable gas around the cannula.
Computational study
Hypothetical low-flow nasal-cannula scenarios varying minute ventilation, inspiratory duty cycle, mean/peak inspiratory flow, nasal capture and reservoir retention.
Modelled hydrogen device flow, source fraction and breathing conditions. Suggested 1–4% inspired-hydrogen scenarios are author-proposed interpretations of earlier literature. · Mean inspiratory flow (MIF) = minute ventilation / inspiratory duty cycle (DC). Without a reservoir, FiH₂ ≈ capture efficiency × H₂ flow / MIF. Reservoir contribution depends on retained expiratory gas and tidal volume. Common scenarios use MIF 30 L/min, DC 0.4, capture 0.85, retention 0.20 and reservoir volume 30–40 mL. Henry-law estimates assume temperature, pressure and equilibrium; they are not measured patient concentrations.
Not applicable; no treatment was administered.
The same device flow produced different modelled inspired fractions when breathing, capture and reservoir assumptions changed. One no-reservoir example uses H₂ flow 600 mL/min, MIF 30 L/min and capture 0.99 to calculate 1.98% FiH₂; an assumed reservoir adds 0.8 percentage points to 2.78%. These are calculations, not patient measurements. Table 7 reports approximately 261, 520 and 1,200 mL/min for 1%, 2% and 4% FiH₂ in one scenario (MIF 30 L/min, DC 0.4, capture 0.85, retention 0.20); other assumptions give wider ranges. Its coverage percentiles are engineering margins, not measured population percentiles. The paper states that the lowest required FiH₂ has not been directly established in human clinical studies and the framework was not prospectively validated.
Author-proposed modelling assumptions and concentration targets are not measured patient exposure, established clinical thresholds or dosing guidance.
What kind of evidence is this?
Computational study
Deterministic respiratory-physiology modelling with targeted literature synthesis; no new experimental or clinical data.
Respiratory health
Inhaled H₂
Computational model output
Calculated inspired-hydrogen fractions and device-flow requirements under specified assumptions.
Methods
Hypothetical low-flow nasal-cannula scenarios varying minute ventilation, inspiratory duty cycle, mean/peak inspiratory flow, nasal capture and reservoir retention.
No people, animals or cultured cells. Scenarios are computational assumptions, not trial participants.
Not applicable; no treatment was administered.
Modelled hydrogen device flow, source fraction and breathing conditions. Suggested 1–4% inspired-hydrogen scenarios are author-proposed interpretations of earlier literature.
Inhalation-delivery model; no hydrogen administered in this study.
Mean inspiratory flow (MIF) = minute ventilation / inspiratory duty cycle (DC). Without a reservoir, FiH₂ ≈ capture efficiency × H₂ flow / MIF. Reservoir contribution depends on retained expiratory gas and tidal volume. Common scenarios use MIF 30 L/min, DC 0.4, capture 0.85, retention 0.20 and reservoir volume 30–40 mL. Henry-law estimates assume temperature, pressure and equilibrium; they are not measured patient concentrations.
Reservoir versus no reservoir, differing capture and retention, duty cycles, inspiratory flows, altitude and source-gas configurations.
Outcomes and reported result
Calculated inspired-hydrogen fractions and device-flow requirements under specified assumptions.
The same device flow produced different modelled inspired fractions when breathing, capture and reservoir assumptions changed. One no-reservoir example uses H₂ flow 600 mL/min, MIF 30 L/min and capture 0.99 to calculate 1.98% FiH₂; an assumed reservoir adds 0.8 percentage points to 2.78%. These are calculations, not patient measurements. Table 7 reports approximately 261, 520 and 1,200 mL/min for 1%, 2% and 4% FiH₂ in one scenario (MIF 30 L/min, DC 0.4, capture 0.85, retention 0.20); other assumptions give wider ranges. Its coverage percentiles are engineering margins, not measured population percentiles. The paper states that the lowest required FiH₂ has not been directly established in human clinical studies and the framework was not prospectively validated.
Complete published full text reviewed. Source contradictions and unreported details remain explicit; no independent reanalysis was performed.
Not applicable to a computational study; no treatment or safety experiment.
A reported association, difference or mechanism is not automatically a clinical benefit.
Limitations and applicability
This is targeted narrative modelling, not a systematic meta-analysis or dosing trial. Capture and reservoir retention were assumed or bounded using oxygen-delivery literature, without an H₂-specific measured distribution. The proposed approximately 1% minimum and 2–4% window are interpretations, not universal consensus thresholds. Henry-law plasma/tissue estimates assume equilibration. Figure 4 panel descriptions and some Table 6 duty-cycle labels are inconsistent; Figure 2 shows a 4% baseline into expiration, which is not an inspired-gas measurement. Average FiH₂ below 4% does not eliminate locally flammable gas around pure-H₂ cannulas. No independent computation, safety-engineering validation or site-calculator audit was performed.
Computational hypothesis or model output, not evidence of clinical efficacy or safety.
JNK reports NIH/NHLBI 4R33HL164373 and 1R01HL175158-01, and NIH/NICHD 1R43HD116647-01.
TWL is Scientific Officer of InhaleH2.com. MS reports advisory fees and research funding from DoctorsMan Co Ltd. JNK reports federal grants. Other authors declare no conflicts.
No single-study GRADE certainty rating is assigned. Read how records and evidence assessments are prepared.
Sources and record status
PMID: 42050516 · DOI: 10.1186/s12931-026-03664-9
Published open-access full text in PMC.
Source verification based on the supplied complete full-text review. This is not independent medical or statistical validation.
Complete main text, figures, tables and funding/conflict disclosures. No independent statistical validation.
Complete published full-text article reviewed, including methods, results, tables, figures, discussion and disclosures.
17 November 2026